亚洲一区二区三区麻豆,日本一区不卡在线观看,蜜桃久久麻豆传媒,午夜成人视频官方版

本站熱搜:氧化鋯氧量分析儀,盤裝式氧量分析儀,電磁流量計(jì),WSS雙金屬溫度計(jì)
  • 技術(shù)文章ARTICLE

    您當(dāng)前的位置:首頁 > 技術(shù)文章 > Common faults and causes of zirconia oxygen analyzers

    Common faults and causes of zirconia oxygen analyzers

    發(fā)布時(shí)間: 2024-12-13  點(diǎn)擊次數(shù): 1741次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


產(chǎn)品中心 Products
在線客服 聯(lián)系方式 二維碼

服務(wù)熱線

0550-2409223

掃一掃,關(guān)注我們

皖公網(wǎng)安備 34118102000850號(hào)

女生和男生交配视频| 成人精品无码色欲| 精品视频666个人主页| 天天操天天舔天天干| 天天干天天操天天射| 尤物在线视频精品国产| 《亚洲视频a》免费在线观看高清完整版 | 岁少女拍性爱视频| 国产在线精品一区二区网站| 国产av在在官方版 | 久久按摩地址| 《99一级特黄色性生活片久久无码》| 成人综合久久精品色婷婷| 无码》电影完整版在线观看| 日本 码无 精 在线 免费| 人片在线观看网站| 色你妹美女动态图| 成人娱乐平台的创新领军者| 《久久香蕉国产精品》高清在线观看地址 | 桃色视频下载| 视频库安卓版| 成人a片小说在线阅读| 全集BD高清完整版| 亚洲精品无码av中文字幕电影| 州区二区三区四区产品分域管理逻辑| 在线字幕翻译| 《性生活黄色毛片》电影在线观看| 院线免费观看| 《日韩a一级》最近中文在线观看| 人妻少妇精品久久久久久久| 彭久洋简历附图| 最新在线防屏蔽国产一区| 性色av免费观看akhss| 网一区二区内容选择要谨慎| 老司机直称好用| 《日韩a级》电影在线观看| 《欧美日韩》电视剧在线观看| 天堂草原电视剧大全| 日韩a一级欧美一级在线播放| 精品国产麻豆婷婷洗澡| 欧美在线视频在线观看h|